Title:Validation and performance assessment of a preclinical SiPM-based SPECT/MRI insert
Creators Name:Carminati, M., Baratelli, F.M., Occhipinti, M., Erlandsson, K., Nagy, K., Nyitrai, Z., Czeller, M., Kühne, A., Niendorf, T., Valtorta, S., Belloli, S., Moresco, R.M., Savi, A., Iadanza, A., Falini, A., Politi, L.S., Cadioli, M., Hutton, B.F. and Fiorini, C.
Abstract:A preclinical insert for small animal simultaneous SPECT and MR imaging, in particular for imaging mouse brains, is presented. It consists of ten static magnetic resonance imaging (MRI)-compatible gamma cameras based on tiles of silicon photomultipliers readout by a multichannel ASIC and coupled to 5 cm × 5 cm CsI(Tl) scintillators and to an MRI-compatible multipinhole collimator. Calibration and image reconstruction algorithm are illustrated. Mutual compatibility is demonstrated along with imaging performance that is comparable with other non-MR micro-SPECT systems: 0.9 mm tomographic spatial resolution across a transverse field of view of 15.6 mm, 12% energy resolution (at 140 keV), and 1105 cps/MBq sensitivity. Experimental results with phantoms (glass capillaries of 290 μm diameter and a mini Derenzo) are presented.
Keywords:Gamma Camera, Multimodal Imaging, Multipinhole Collimator, Silicon Photomultipliers (SiPMs)
Source:IEEE Transactions on Radiation and Plasma Medical Sciences
ISSN:2469-7311
Publisher:IEEE
Volume:3
Number:4
Page Range:483-490
Date:July 2019
Official Publication:https://doi.org/10.1109/TRPMS.2019.2893377