Item Type: | Article |
---|---|
Title: | Corrections of myocardial tissue sodium concentration measurements in human cardiac (23)Na MRI at 7 Tesla |
Creators Name: | Lott, J., Platt, T., Niesporek, S.C., Paech, D., Behl, N.G.R., Niendorf, T., Bachert, P., Ladd, M.E. and Nagel, A.M. |
Abstract: | PURPOSE: To quantify the tissue sodium concentration (TSC) in cardiac (23)Na MRI. To evaluate the influence of different correction methods on the measured myocardial TSC. METHODS: (23)Na MRI of four healthy subjects was conducted at a whole-body 7T MRI system using an oval-shaped (23)Na birdcage coil. Data acquisition was performed with a density-adapted 3D radial pulse sequence using a golden angle projection scheme. (1)H MRI data were acquired at a 3T MRI system to generate a myocardial mask. Retrospective cardiac and respiratory gating were used to reconstruct (23)Na MRI data in the diastolic phase and exhaled state. B(0) and B(1) inhomogeneity and partial volume (PV) effects were corrected. Relaxation times and TSC of ex vivo blood samples and calf muscle were determined. These values were used in the PV correction to estimate myocardial TSC, which was compared with the measured TSC of calf muscle. RESULTS: Without any correction the measured myocardial TSC was (54 ± 5) mM. The applied correction methods reduced these values by (48 ± 5)% to (29 ± 3) mM, where PV correction had the largest effect (reduction of (34 ± 1)%). Respiratory and cardiac motion gating decreased the concentrations by (11 ± 1)%. With the applied setup, the corrections of B(0) and B(1) inhomogeneity (reduction of (3 ± 2)%) had negligible influences on TSC values. The resulting myocardial TSC was approximately 1.4-fold higher than the measured TSC of calf muscle tissue of the same healthy subjects ((20 ± 3) mM). CONCLUSION: For quantitative human cardiac (23)Na MRI several corrections are needed and ranked for our setup: PV correction, respiratory and cardiac gating, correction for B(1) inhomogeneity effects. |
Keywords: | 7 Tesla, Oval-Shaped Body Coil, Partial Volume Correction, Quantitative Sodium ((23)Na) Cardiac MRI, Retrospective Cardiac and Respiratory Gating |
Source: | Magnetic Resonance in Medicine |
ISSN: | 0740-3194 |
Publisher: | Wiley |
Volume: | 82 |
Number: | 1 |
Page Range: | 159-173 |
Date: | July 2019 |
Official Publication: | https://doi.org/10.1002/mrm.27703 |
PubMed: | View item in PubMed |
Repository Staff Only: item control page