Helmholtz Gemeinschaft

Search
Browse
Statistics
Feeds

A simulation-based phantom model for generating synthetic mitral valve image data-application to MRI acquisition planning

[img]
Preview
PDF (Original Article) - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
4MB
[img] Other (Supplementary Materials)
1MB

Item Type:Article
Title:A simulation-based phantom model for generating synthetic mitral valve image data-application to MRI acquisition planning
Creators Name:Manini, C. and Nemchyna, O. and Akansel, S. and Walczak, L. and Tautz, L. and Kolbitsch, C. and Falk, V. and Sündermann, S. and Kühne, T. and Schulz-Menger, J. and Hennemuth, A.
Abstract:PURPOSE: Numerical phantom methods are widely used in the development of medical imaging methods. They enable quantitative evaluation and direct comparison with controlled and known ground truth information. Cardiac magnetic resonance has the potential for a comprehensive evaluation of the mitral valve (MV). The goal of this work is the development of a numerical simulation framework that supports the investigation of MRI imaging strategies for the mitral valve. METHODS: We present a pipeline for synthetic image generation based on the combination of individual anatomical 3D models with a position-based dynamics simulation of the mitral valve closure. The corresponding images are generated using modality-specific intensity models and spatiotemporal sampling concepts. We test the applicability in the context of MRI imaging strategies for the assessment of the mitral valve. Synthetic images are generated with different strategies regarding image orientation (SAX and rLAX) and spatial sampling density. RESULTS: The suitability of the imaging strategy is evaluated by comparing MV segmentations against ground truth annotations. The generated synthetic images were compared to ones acquired with similar parameters, and the result is promising. The quantitative analysis of annotation results suggests that the rLAX sampling strategy is preferable for MV assessment, reaching accuracy values that are comparable to or even outperform literature values. CONCLUSION: The proposed approach provides a valuable tool for the evaluation and optimization of cardiac valve image acquisition. Its application to the use case identifies the radial image sampling strategy as the most suitable for MV assessment through MRI.
Keywords:Image Simulation, Mitral Valve, Modeling, Segmentation, Cardiac Phantom, Magnetic Resonance Imaging
Source:International Journal of Computer Assisted Radiology and Surgery
ISSN:1861-6410
Publisher:Springer
Volume:19
Number:3
Page Range:553-569
Date:March 2024
Official Publication:https://doi.org/10.1007/s11548-023-03012-y
PubMed:View item in PubMed

Repository Staff Only: item control page

Downloads

Downloads per month over past year

Open Access
MDC Library