Simultaneous SPECT/MR imaging with a SiPM-based preclinical insert

Item Type:Conference or Workshop Item
Title:Simultaneous SPECT/MR imaging with a SiPM-based preclinical insert
Creators Name:Carminati, M. and Baratelli, F.M. and Massara, M. and Occhipinti, M. and Nagy, K. and Nyitrai, Z. and Czeller, M. and Kühne, A. and Niendorf, T. and Valtorta, S. and Belloli, S. and Moresco, R.M. and Savi, A. and Iadanza, A. and Falini, A. and Politi, L.S. and Cadioli, M. and Fiorini, C.
Abstract:The results of a systematic and quantitative experimental characterization of simultaneous operation of a preclinical SPECT insert inside an unmodified standard MR scanner (Philips 3T Achieva) are presented. The static gantry (20 cm diameter, 15 mm field of view) is composed of 10 MRI-compatible gamma cameras based on 8-mm thick CsI(Tl) scintillator (50 mm × 50mm) coupled with arrays of SiPMs. Mutual compatibility is achieved by proper design of the SPECT system: shielding, choice of materials, design of the analog boards. The measured energy resolution is 13% at 122 keV after correction of SiPM gain variability and local crystal luminosity. The intrinsic spatial resolution, after planar image statistical reconstruction by means of maximum likelihood, is 1mm, while the extrinsic resolution is below 0.9 mm across the entire transaxial FoV, measured with capillaries of 290 pm diameter. With the custom-designed 3 × 3 multi-pinhole collimator (whose minimal perturbation of the magnetic field can be reduced below 1ppm thanks to shimming), the measured sensitivity with 99m Tc is 1100 cps/MBq. The MR image is not distorted by the operation of the SPECT, as well as SPECT performances are no degraded by MRI (T2W and gradient sequences). Simultaneous imaging of different phantoms (capillaries, Derenzo hot rods of 0.85-1.7 mm diameter) as well as a mouse brain are reported.
Keywords:Single Photon Emission Computed Tomography, Magnetic Resonance Imaging, Collimators, Spatial Resolution, Energy Resolution, Instruments
Source:2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC)
Title of Book:2018 IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC)
ISSN:2577-0829
ISBN:978-1-5386-8494-8
Publisher:IEEE
Page Range:1-3
Date:5 September 2019
Official Publication:https://doi.org/10.1109/NSSMIC.2018.8824631