Helmholtz Gemeinschaft

Search
Browse
Statistics
Feeds

Impact of sequence type and field strength (1.5, 3, and 7T) on 4D flow MRI hemodynamic aortic parameters in healthy volunteers

[img]
Preview
PDF (Original Article) - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
1MB
[img] Other (Supplementary Material)
7MB

Item Type:Article
Title:Impact of sequence type and field strength (1.5, 3, and 7T) on 4D flow MRI hemodynamic aortic parameters in healthy volunteers
Creators Name:Wiesemann, S. and Schmitter, S. and Demir, A. and Prothmann, M. and Schwenke, C. and Chawla, A. and von Knobelsdorff-Brenkenhoff, F. and Greiser, A. and Jin, N. and Bollache, E. and Markl, M. and Schulz-Menger, J.
Abstract:PURPOSE: 4D flow magnetic resonance imaging (4D‐MRI) allows time‐resolved visualization of blood flow patterns, quantification of volumes, velocities, and advanced parameters, such as wall shear stress (WSS). As 4D‐MRI enters the clinical arena, standardization and awareness of confounders are important. Our aim was to evaluate the equivalence of 4D flow‐derived aortic hemodynamics in healthy volunteers using different sequences and field strengths. METHODS: 4D‐MRI was acquired in 10 healthy volunteers at 1.5T using three different prototype sequences, at 3T and at 7T (Siemens Healthineers). After evaluation of diagnostic quality in three segments (ascending‐, descending aorta, aortic arch), peak velocity, flow volumes, and WSS were investigated. Equivalence limits for comparison of field strengths/sequences were based on the limits of Bland‐Altman analyses of the intraobserver variability. RESULTS: Non‐diagnostic quality was found in 10/144 segments, 9/10 were obtained at 7T. Apart for the comparison of forward flow between sequence 1 and 3, the differences in measurements between field strengths/sequences exceeded the range of agreement. Significant differences were found between field strengths/sequences for forward flow (1.5T vs. 3T, 3T vs. 7T, sequence 1 vs. 3, 2 vs. 3 [P < .001]), WSS (1.5T vs. 3T [P < .05], sequence 1 vs. 2, 1 vs. 3, 2 vs. 3 [P < .001]), and peak velocity (1.5T vs. 7T, sequence 1 vs. 3 [P > .001]). All parameters at all field strengths/with all sequences correlated moderately to strongly (r ≥ 0.5). CONCLUSION: Data from all sequences could be acquired and resulting images showed sufficient quality for further analysis. However, the variability of the measurements of peak velocity, flow volumes, and WSS was higher when comparing field strengths/sequences as the equivalence limits defined by the intraobserver assessments.
Keywords:4D Flow, 7T, Aorta, Cardiovascular Magnetic Resonance Imaging, Non-Invasive Hemodynamics, Standardization
Source:Magnetic Resonance in Medicine
ISSN:0740-3194
Publisher:Wiley
Volume:85
Number:2
Page Range:721-733
Date:February 2021
Additional Information:Copyright © 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.
Official Publication:https://doi.org/10.1002/mrm.28450
PubMed:View item in PubMed

Repository Staff Only: item control page

Downloads

Downloads per month over past year

Open Access
MDC Library