A multi-vendor, multi-center study on reproducibility and comparability of fast strain-encoded cardiovascular magnetic resonance imaging

[img]
Preview
PDF (Original Article) - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
1MB

Item Type:Article
Title:A multi-vendor, multi-center study on reproducibility and comparability of fast strain-encoded cardiovascular magnetic resonance imaging
Creators Name:Erley, J. and Zieschang, V. and Lapinskas, T. and Demir, A. and Wiesemann, S. and Haass, M. and Osman, N.F. and Simonetti, O.P. and Liu, Y. and Patel, A.R. and Mor-Avi, V. and Unal, O. and Johnson, K.M. and Pieske, B. and Hansmann, J. and Schulz-Menger, J. and Kelle, S.
Abstract:Myocardial strain is a convenient parameter to quantify left ventricular (LV) function. Fast strain-encoding (fSENC) enables the acquisition of cardiovascular magnetic resonance images for strain-measurement within a few heartbeats during free-breathing. It is necessary to analyze inter-vendor agreement of techniques to determine strain, such as fSENC, in order to compare existing studies and plan multi-center studies. Therefore, the aim of this study was to investigate inter-vendor agreement and test-retest reproducibility of fSENC for three major MRI-vendors. fSENC-images were acquired three times in the same group of 15 healthy volunteers using 3 Tesla scanners from three different vendors: at the German Heart Institute Berlin, the Charité University Medicine Berlin-Campus Buch and the Theresien-Hospital Mannheim. Volunteers were scanned using the same imaging protocol composed of two fSENC-acquisitions, a 15-min break and another two fSENC-acquisitions. LV global longitudinal and circumferential strain (GLS, GCS) were analyzed by a trained observer (Myostrain 5.0, Myocardial Solutions) and for nine volunteers repeatedly by another observer. Inter-vendor agreement was determined using Bland-Altman analysis. Test-retest reproducibility and intra- and inter-observer reproducibility were analyzed using intraclass correlation coefficient (ICC) and coefficients of variation (CoV). Inter-vendor agreement between all three sites was good for GLS and GCS, with biases of 0.01-1.88%. Test-retest reproducibility of scans before and after the break was high, shown by ICC- and CoV values of 0.63-0.97 and 3-9% for GLS and 0.69-0.82 and 4-7% for GCS, respectively. Intra- and inter-observer reproducibility were excellent for both parameters (ICC of 0.77-0.99, CoV of 2-5%). This trial demonstrates good inter-vendor agreement and test-retest reproducibility of GLS and GCS measurements, acquired at three different scanners from three different vendors using fSENC. The results indicate that it is necessary to account for a possible bias (< 2%) when comparing strain measurements of different scanners. Technical differences between scanners, which impact inter-vendor agreement, should be further analyzed and minimized.DRKS Registration Number: 00013253. Universal Trial Number (UTN): U1111-1207-5874.
Keywords:Strain, fSENC, Agreement, Reproducibility, CMR, Cardiac, Magnetic Resonance
Source:International Journal of Cardiovascular Imaging
ISSN:1569-5794
Publisher:Springer
Volume:36
Number:5
Page Range:899-911
Date:May 2020
Official Publication:https://doi.org/10.1007/s10554-020-01775-y
PubMed:View item in PubMed

Repository Staff Only: item control page

Downloads

Downloads per month over past year

Open Access
MDC Library