Helmholtz Gemeinschaft

Search
Browse
Statistics
Feeds

Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements

[img] PDF - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
1MB

Item Type:Article
Title:Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
Creators Name:Gatehouse, P.D. and Rolf, M.P. and Graves, M.J. and Hofman, M.B. and Totman, J. and Werner, B. and Quest, R.A. and Liu, Y. and von Spiczak, J. and Dieringer, M. and Firmin, D.N. and van Rossum, A. and Lombardi, M. and Schwitter, J. and Schulz-Menger, J. and Kilner, P.J.
Abstract:Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. We studied the severity of uncorrected velocity offset errors across sites and CMR systems.
Keywords:Aorta, Artifacts, Blood Flow Velocity, Cardiac Output, Gelatin, Computer-Assisted Image Interpretation, Cine Magnetic Resonance Imaging, Materials Testing, Mitral Valve Insufficiency, Imaging Phantoms, Predictive Value of Tests, Pulmonary Artery, Reproducibility of Results, Respiratory Mechanics
Source:Journal of Cardiovascular Magnetic Resonance
ISSN:1097-6647
Publisher:BioMed Central
Volume:12
Page Range:5
Date:14 January 2010
Official Publication:https://doi.org/10.1186/1532-429X-12-5
PubMed:View item in PubMed

Repository Staff Only: item control page