Helmholtz Gemeinschaft

Search
Browse
Statistics
Feeds

APOSTEL 2.0 recommendations for reporting quantitative optical coherence tomography studies

[thumbnail of Original Article]
Preview
PDF (Original Article) - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
495kB

Item Type:Article
Title:APOSTEL 2.0 recommendations for reporting quantitative optical coherence tomography studies
Creators Name:Aytulun, A., Cruz-Herranz, A., Aktas, O., Balcer, L.J., Balk, L., Barboni, P., Blanco, A.A., Calabresi, P.A., Costello, F., Sanchez-Dalmau, B., DeBuc, D.C., Feltgen, N., Finger, R.P., Frederiksen, J.L., Frohman, E., Frohman, T., Garway-Heath, D., Gabilondo, I., Graves, J.S., Green, A.J., Hartung, H.P., Havla, J., Holz, F.G., Imitola, J., Kenney, R., Klistorner, A., Knier, B., Korn, T., Kolbe, S., Krämer, J., Lagrèze, W.A., Leocani, L., Maier, O., Martínez-Lapiscina, E.H., Meuth, S., Outteryck, O., Paul, F., Petzold, A., Pihl-Jensen, G., Preiningerova, J.L., Rebolleda, G., Ringelstein, M., Saidha, S., Schippling, S., Schuman, J.S., Sergott, R.C., Toosy, A., Villoslada, P., Wolf, S., Yeh, E.A., Yu-Wai-Man, P., Zimmermann, H.G., Brandt, A.U. and Albrecht, P.
Abstract:OBJECTIVE: To update the consensus recommendations for reporting of quantitative optical coherence tomography (OCT) study results, thus revising the previously published Advised Protocol for OCT Study Terminology and Elements (APOSTEL) recommendations. METHODS: To identify studies reporting quantitative OCT results, we performed a PubMed search for the terms “quantitative” and “optical coherence tomography” from 2015 to 2017. Corresponding authors of the identified publications were invited to provide feedback on the initial APOSTEL recommendations via online surveys following the principle of a modified Delphi method. The results were evaluated and discussed by a panel of experts, and changes to the initial recommendations were proposed. A final survey was recirculated among the corresponding authors to obtain a majority vote on the proposed changes. RESULTS: One hundred sixteen authors participated in the surveys, resulting in 15 suggestions, of which 12 were finally accepted and incorporated into an updated 9-point-checklist. We harmonized the nomenclature of the outer retinal layers, added the exact area of measurement to the description of volume scans; we suggested reporting device-specific features. We advised to address potential bias in manual segmentation or manual correction of segmentation errors. References to specific reporting guidelines and room light conditions were removed. The participants’ consensus with the recommendations increased from 80% for the previous APOSTEL version to greater than 90%. CONCLUSIONS: The modified Delphi method resulted in an expert-led guideline (evidence class III, GRADE criteria) concerning study protocol, acquisition device, acquisition settings, scanning protocol, fundoscopic imaging, post-acquisition data selection, post-acquisition analysis, nomenclature and abbreviations, and statistical approach. It will still be essential to update these recommendations to new research and practices regularly.
Keywords:Consensus, Delphi Technique, Ophthalmology, Optical Coherence Tomography, Research Design, Retinal Diseases
Source:Neurology
ISSN:0028-3878
Publisher:American Academy of Neurology
Volume:97
Number:2
Page Range:68-79