Glass-capillary collimator for distance compensation and partial monochromatization at rotating-anode X-ray generators

[img] PDF - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
1MB

Item Type:Article
Title:Glass-capillary collimator for distance compensation and partial monochromatization at rotating-anode X-ray generators
Creators Name:Mueller, J.J. and Gorny, H.E. and Schmalz, J. and Heinemann, U.
Abstract:Access to the beam ports of rotating-anode X-ray generators is often obstructed by direct-coupled or belt-driven target drives. The construction of an easily adjustable stable glass-capillary collimator is described, which renders possible the unrestricted use of beam ports of these generators. Transmitted intensity and monochromaticity of the primary beam are sufficient for precession photographs of proteins after additional 20 mu m Ni filtering as demonstrated by a precession photograph of hen egg lysozyme. The straight capillary collimator is now a routinely usable low-cost device for each X-ray laboratory.
Keywords:Spectrometer
Source:Journal of Applied Crystallography
ISSN:0021-8898
Publisher:International Union of Crystallography
Volume:28
Number:Pt 6
Page Range:853-855
Date:1 December 1995
Official Publication:https://doi.org/10.1107/S0021889895009708